This instrument is designed and manufactured according to the requirements of T 0604-2011 "Asphalt Penetration Test" in JTG E20-2011 "Test Methods for Highway Engineering Asphalt and Asphalt Mixtures", the standard of the Ministry of Transport of the People's Republic of China. It is suitable for testing the penetration of road petroleum asphalt, modified asphalt, and the residue after evaporation of liquid petroleum asphalt or emulsified asphalt.
Complies with JJG(Transportation)067-2015 "Measurement Method of Asphalt Penetration", GB4509-84 "Measurement Method of Petroleum Asphalt Penetration", and TO60-93 "Asphalt Penetration Test"
Main Technical Features
- The lifting mechanism is novel in design, and the fine-tuning structure makes the up and down adjustment easy and smooth.
2. Equipped with a special low-temperature circulating constant-temperature water bath to provide circulating constant-temperature water to the glass container on the penetration tester's work platform, creating the required, pre-set, constant low-temperature environment for the sample being tested.
3. Adopts a novel digital display, stable data, reliable operation, and accurate measurement results.
- Advanced digital displacement sensor ensures accurate data measurement.
- LED cold light source illumination ensures accurate water temperature.
The biggest features of this instrument are: equipped with a low-temperature constant-temperature water bath to meet the needs of low-temperature testing; advanced mechanical design for convenient lifting and adjustment; digital temperature control meter for accurate temperature control; high-precision digital penetration test meter for reliable measurement results.
Main Technical Specifications
- Constant Temperature Water Bath Parameters
⑴ Power supply: AC (20±10%) V; 50Hz.
⑵ Bath capacity: 5.5L.
⑶ Temperature range: (5.0~100.0)℃.
⑷ Temperature control accuracy: ±0.1℃.
⑸ Circulating water volume: 5L/min.
⑹ Water pump head: 1m.
⑺ Total power consumption: No more than 1800W.
⑻ External dimensions: 65×41×55cm (length×width×height)
(9)Weight: 29kg
2. Penetration Tester Parameters
⑴ Time control device: 5 seconds, 10 seconds, 30 seconds, and 60 seconds can be selected separately, with a time error of less than ±0.1 seconds.
⑵ Penetration depth device: Uses a high-precision digital display, convenient, fast, and intuitive to use, and can be started at any position.
⑶ Resolution 0.01㎜;
⑷ Penetration accuracy: ±1 penetration.
⑸ Maximum penetration depth: 500 penetration.
⑹ Standard needle: 50㎜, 2.5g±0.05g (basic configuration).
⑺ Penetration external dimensions: 37 x25x48cm (length×width×height)